IV and Liquid Filters: Epi-Max

Epi-Max

Dimensions

Width: 30.0 mm
Length: 62.8 mm
Depth: 7.2 mm

Weight

7.0 gr

Effective Filtration Area

5.0 cm2

Efficiency of Filtration

100 %

Inlet/Outlet Connectors

Inlet: FLL (Female Luer Lock)
Outlet: RMLL (Rotating Male Luer Lock)

Materials of construction

Housing: Clear Modified Acrylic
Filter media: Hydrophilic HI-FLO PES membrane 0,2 - 1,2µm
PES membrane 5,0 µm

Minimum Water Flow Rate

0.2pos μm : >10 ml/min@80 cm (31.5 in) water head pres.
0.2 μm : >15 ml/min@80 cm (31.5 in) water head pressure
1.2 μm : >90 ml/min@80 cm (31.5 in) water head pressure
5.0 μm : >95 ml/min@80 cm (31.5 in) water head pressure

Application

ANESTHESIA
INFUSION

Minimum Water Bubble Point

0.2/0.2pos μm: 3.7-4,8 bar
1.2 μm: 0.7-1,0 bar
5.0 μm: 0.15-0.3 bar

Biosafety

Materials comply with USP class VI-121°C test and ISO 10993

Ref. Standard

ISO 8536-4

Assembly Method

connectors

Pyrogenicity

<0,25 EU/ml

Operating Pressure

Housing resistance - Max burst pressure: ramped pressure 8.0 bar

Maximum Operating Temperature

5 - 40 °C

Sterilization

EtO or Gamma (25kGy)

Bacterial Retention

0.2 µm: brevundimonas diminuta
0.2 µm positive: endotoxin
1.2 µm: candida albicans

 

Ordering Information:

Product Code
Description
RS065DCYRH002A00
Epi-Max HI-FLO 0.2 micron logo FLL IN/RMLL OUT
RS065DCYRH012A00
Epi-Max HI-FLO 1.2 micron logo FLL IN/RMLL OUT
RS065DCYRA050A00
Epi-Max HI-FLO 5.0 micron logo FLL IN/RMLL OUT
RS065DCYCA002A00
Epi-Max HI-FLO 0.2 micron positive logo FLL IN/RMLL OUT

Packaging:

Dimension
Weight
Quantity / Box
60x40x25 cm
14 kg
2,000 units

Notes

Epi-Max – Epidural Filters in Amber Modified Acrylic for photosensitive drugs are available upon request